This standard operating procedure (S.O.P.) is intended for all nursing staff required to set-up (Manual) Acute Peritoneal Dialysis using the Dialy-Nate® closed peritoneal dialysis set within the Paediatric Critical Care Unit at the Royal Hospital for Children (Glasgow).
All nursing staff involved in caring for infants or children in the Paediatric Critical care unit requiring manual acute peritoneal dialysis, should be familiar with this S.O.P.
Renal replacement therapy is an important supportive measure used in paediatric intensive care for the management of Acute kidney injury (AKI). Whilst extracorporeal renal replacement therapies, such as haemofiltration/haemodiafiltration are the more commonly used modality, peritoneal dialysis is still a useful, safe and inexpensive procedure. 1,2,3,4
Advantages:
P/D in paediatrics has several advantages over extracorporeal therapies and includes the following: the peritoneal surface area in paediatric patients is greater than that of adults, thus P/D can result in a more efficient solute clearance than that of P/D in adult patients. 5 P/D is a dynamic dialysis and more physiological process, resulting in less pro-inflammatory consequences than that of Hemodialysis, for example. This continuous process, allows gradual solute and fluid removal, is safe, and is well tolerated cardiovascularly in the critically ill child. 1,4 Finally, P/D is technically far simpler than using extracorporal renal replacement therapies, as it requires minmal infrastructure support and can be perfomed by the majority of PICU bedside nurses. 2,4
Disadvantages:
There are limitations to using acute P/D in paediatric intensive care, including the following situations: an intact paritioneal cavity is essential for acute P/D therefore it is not a suitable renal support for those children with conditions such as recent abdominal surgeries, diaphragmatic hernia or parlaytic ileus. 5 As ultrafiltration and solute removal can be unpredicatable in acute P/D, it is also not the ideal choice for those conditions where fluid balance must be precise and controlled, such as multi-organ failure and shock. Children in these conditions will have poor tolerance of increased peritoneal/abdominal volume from the P/D ‘dwell’ volume and decreased splanchnic perfusion. Thus, effective and rapid correction of any life-threatening fluid and electrolyte problems may be difficult with acute P/D. Finally, there are mechnical complications associated with using acute P/D, including catheter leakage and catheter obstruction which can result in reduction in efficiency of dialysis. 1,6
Some other associated problems can include risk of hyperglycaemia (from dialysis fluid used) and hypothermia if fluid not warmed.
The decision about what dialysis modality should be used is based upon the child’s clinical condition, the availability of necessary resources and the expertise and skills of the local staff.
Dialy-Nate® P/D Set (x1) (E.g. 400537) 2.5 or 5 litre bag prescribed dialysis solution (E.g. Physioneal 1.36% or 2.27%) |
Disposable apron Gloves Sterile drape Trolley IV pole / stand Chlorhexidine 2% & Isopropyl Alcohol 70% cleaning swab (E.g. Clinell® wipe) Connector shield with povidone-iodine (Baxter SPC4211)*** Extension tubing (Baxter R5C4482) Minicap (Baxter SPC4466)
(***note - due to iodine content check use for infants <1yr)
|
PROCEDURE |
RATIONALE |
Provide age appropriate explanation of the procedure. |
To ensure that the child understands the procedure and to avoid undue distress.
|
Positioning infant/child Ensure unobstructed access to PD catheter connection Check bloods – glucose, Monitoring – minimum: ECG, SpO2, BP & +/- |
|
PROCEDURE |
RATIONALE |
Gather all equipment required and take to bedside Clean trolley |
|
Wash hands thoroughly with appropriate antibacterial skin cleanser then don disposable gloves.7
Open drape onto trolley |
In order to minimise the risk of cross- infection. |
Open outer packaging of Dialy-nate® system set and open onto trolley top (keep packaging for ref no) |
|
Open cleaning wipe, connection shield*** and extension tubing on to drape
(*** note-due to iodine content check use for infants <1yr) |
|
Open dialysis solution bag onto trolley top.
If the solution is in two sections (E.g. Physioneal) (Fig.), break frangible connector inside (Fig..) between ‘top’ and ‘bottom’ bag, and ensure fluid flows into ‘bottom’ bag and solution is well mixed. |
|
Hang dialysis solution bag on pole |
PROCEDURE |
RATIONALE |
Connect dialysis solution bag to the Dialy-Nate® set, using one of the luer connectors. |
|
Clamp all 3 luer-connector lines
Close the roller clamp (roller clamp is below buretrol chamber and filter) |
|
Check PD extension set attached to ‘Patient’ line on system. |
|
Ensure warming coil loop in circuit is positioned to allow priming of system and air removal |
|
Hang drainage bag / collection system (on trolley) ensuring it is below the height of the buretrol and tubing |
|
Break the frangible inside connector on the bottom of the dialysis solution bag and then open the clamp on that connector line only. |
|
Keeping the buretrol upright, allow 60-80ml of dialysis solution to drain into the buretrol, then clamp that connector. |
|
Keeping the patient line of the Dialy-nate® set on the drape, open roller clamp below buretrol and open stopcock to ‘open’ on patient line and prime to the end of the patient line (including PD extension set). |
|
Once this line is primed, turn the stopcock to ‘closed’ on the patient line and ‘open’ to the drain line. |
|
Let the remainder of the fluid prime through – but STOP (i.e. close the stopcock at 2 o’clock) when volume in buretrol is 10ml. This is to prevent air entering the circuit. |
|
Once primed, then label all tubing on the system: - ‘Fill’ - ‘Patient’ and - ‘Drain’. |
|
Make sure buretrol is upright on pole and that all tubing in/out of P/D catheter is not kinked or looped. |
PROCEDURE |
RATIONALE |
Using Clinell® or similar wipe, clean the connector at end of tubing on child’s PD catheter.
Allow to dry. |
|
Connect white end of PD extension set and connect to patient PD cannula.
Take tubing labelled ‘Patient’, and connect to PD extension set. |
|
Wrap connection shield around blue end of PD extension set.
*Renal staff handy hint: you can connect the Connection shield to the primed system 1st - then connect to the patient PD extension set. It makes it easier to get ‘snug’ fit. ONLY do this if you confident you have a secure connection. |
* |
PROCEDURE |
RATIONALE |
Ensure primed system and lines are secured to cot/bed side and not obstructed |
|
Once connected, open the stopcock to the ‘drain’ line and let drain for the prescribed amount of time. Set timer for this – record this initial drainage. |
|
Clamp ‘drain’ line by turning stopcock to 2 o’clock position (this switches off drain, fill & patient lines) |
|
P/D cycle: a. Fill the buretrol with the desired volume prescribed for child PLUS 10ml. E.g. dialysis solution fill volume: 40ml PLUS 10ml = total 50ml in buretrol.
b. Once buretrol filled to prescribed volume (+10ml) - Clamp dialysis solution bag to buretrol |
|
c. Then: - open stopcock to ‘patient’ line - unclamp roller clamp (below buretrol and filter) - let fluid in the buretrol flow into the P/D catheter (N.B. this may be slow) - once level in buretrol at 10ml* – CLAMP roller clamp *do not let buretrol empty as air will then pass down system |
|
d. Close the stopcock to the ‘patient’ and all lines (i.e. 2 o’clock).
e. Start timer for desired dwell time (E.g. 30 minutes) |
|
f. Once desired dwell time complete, open the stopcock to the ‘drain’ line.
g. Again set timer to prescribed ‘drain out’ time (E.g. 10 minutes or 15 minutes). |
|
h. At the end of this drain out time, close the stopcock to the ‘drain’ line. (2 o’clock position) |
|
i. Measure amount of drainage in the urimeter, then once amount noted, empty this into the larger collection bag.
j. Document amount: This should be the difference between volume of fluid in & volume of drainage out E.g. 40 ml IN / 52 ml OUT = 12ml drained
Repeat steps a to j for the prescribed time for dialysis. |
|
PROCEDURE |
RATIONALE |
Wash hands and don apron and non-sterile gloves.
Ensure stopcock and all clamps are closed switched off on ‘fill’, ‘patient’ and ‘drain’ lines (i.e. @ 2 o’cock position). |
|
Remove connection shield cap and disconnect Dialy-nate® system from patient P/D catheter connection. |
|
Clean P/D connection with Clinell swab and attach minicap to blue end of PD extension set. |
|
Remove bag of dialysis solution from Dialy-nate® system and clamp/cap off if any solution remaining. Place on trolley top. Ensure all drainage volumes checked and recorded then place system on trolley also. Take both bag & drainage system to sluice and empty drainage into sluice then dispose of bag/system as clinical waste. |
Acute P/D using Dialy-nate® system in PICU
Gather all equipment & wash hands . Open Dialy-nate® P/D pack & dialysis fluid on to trolley.
Priming system
11 & 12. Open stopcock to ‘patient’ line and prime system (including prime to PD extension)
Performing manual acute peritoneal dialysis
Before initiation of very 1st PD:
1st cycle and ongoing PD cycles:
Fill burette to prescribed P/D volume E.g. 40ml (+10ml) then clamp at bag & above burette
Set timer for prescribed ‘dwell’ time for PD cycle.
Set timer for prescribed drain out time in cycle.
Note volume drained and document. Repeat timed cycles as prescribed.
Discontinuing Acute Peritoneal Dialysis
Wash hands and don apron and non-sterile gloves.
26.Remove connection shield cap, disconnect Dialy-nate® system from patient P/D catheter connection.
Remove all waste and document volumes
Last reviewed: 30 April 2018
Next review: 30 April 2020
Author(s): J Grady